Monnit Mine Java
|
Parameters |
Description |
Sensor sens |
the Sensor object to be updated |
Double heartBeat |
How often the sensor communicates with the gateway if no activity recorded. |
Integer windowFunction |
No Window = 0 (Best frequency resolution), Hanning Window = 1 (Good frequency resolution and amplitude accuracy), Flattop Window = 2 (Best amplitude accuracy) |
Integer assessmentInterval |
Determines how often the sensor takes a measurement . for example, if the Heartbeat is 10 minutes and the Assessment interval is 1 minute, the sensor will make a measurement every minute, but only report every 10 minutes. |
Integer vibrationIntensityThreshold |
This value determines the g-forces required to wake up the sensor to take a measurement. if the g-forces are not greater than this, the sensor will remain asleep. Values range from 1 to 127, each unit is .063 g-forces. |
Integer failedTransmissionsBeforeLinkMode |
The number of transmissions the sensor sends without response from a gateway before it goes to battery saving link mode. In link mode the sensor will scan for a new gateway and if not found will enter battery saving sleep mode for 2 hours before trying to scan again. Lower number will allow sensors to find new gateways with fewer missed readings. Higher numbers will cause the sensor to hold it setting in a noisy RF environment better. (Zero will cause the sensor to never enter link mode. to find a new gateway channel the battery will have to be cycled out of the sensor.) |
Allows for the update of the specific type of sensor.